Dokument: Differences in the swallowing process of newborns and healthy preterm infants: first results with a non-invasive bioimpedance and electromyography measurement system

Titel:Differences in the swallowing process of newborns and healthy preterm infants: first results with a non-invasive bioimpedance and electromyography measurement system
URL für Lesezeichen:https://docserv.uni-duesseldorf.de/servlets/DocumentServlet?id=68910
URN (NBN):urn:nbn:de:hbz:061-20250304-121345-5
Kollektion:Publikationen
Sprache:Englisch
Dokumententyp:Wissenschaftliche Texte » Artikel, Aufsatz
Medientyp:Text
Autoren: Hübl, Nicole [Autor]
Riebold, Benjamin [Autor]
Schramm, Dirk [Autor]
Seidl, Rainer O. [Autor]
Dateien:
[Dateien anzeigen]Adobe PDF
[Details]1,27 MB in einer Datei
[ZIP-Datei erzeugen]
Dateien vom 04.03.2025 / geändert 04.03.2025
Stichwörter:Breathing, Aspiration, Bioimpedance, Preterm infant, Swallowing, Neonate
Beschreibung:Purpose

Preterm infants (PI) have difficulty coordinating sucking, swallowing and breathing, and there is a risk of aspiration. The causes of this are not yet sufficiently understood. The aim of this study was to test a novel measurement device to measure breathing and pharyngeal processes involved in swallowing externally in everyday life to identify possible differences in neonates (NB) and PI.
Methods

Forty healthy NB were studied at 4–8 weeks of age (mean: 6.7 weeks) and 20 healthy PI (mean gestational age 30.5 weeks) at postmenstrual age (PMA) 34/35 weeks (mean PMA 35.1 weeks) during a single feeding. Surface electrodes were used to measure bioimpedance and electromyography reflecting swallow-related changes in the pharynx and muscle activation of the tongue and submental muscles. A respiratory belt was combined with recording of the depth of chest movements and the occurrence of pauses in breathing.
Results

Velocity and extent of pharyngeal closure did not differ significantly across the feeding period (velocity: p=0.09, closure: p=0.17), but during the first two suck–swallow bursts PI had greater velocity (p<0.001*) and extent of pharyngeal closure (p=0.004*) than NB. The duration of swallowing phases was significantly longer in PIs (p<0.001*), their muscle activation decreased faster (p<0.001*), and they had more pauses in breathing than NBs.
Conclusions

The novel measurement device allowed, for the first time in everyday life, the measurement of factors influencing swallowing and breath–swallow coordination in NBs and PIs. PIs showed differences from NBs most likely due to differences in muscle strength and condition.
Rechtliche Vermerke:Originalveröffentlichung:
Hübl, N., Riebold, B., Schramm, D., & Seidl, R. O. (2023). Differences in the swallowing process of newborns and healthy preterm infants: first results with a non-invasive bioimpedance and electromyography measurement system. European Archives of Oto-Rhino-Laryngology and Head & Neck, 281(2), 843–854. https://doi.org/10.1007/s00405-023-08344-8
Lizenz:Creative Commons Lizenzvertrag
Dieses Werk ist lizenziert unter einer Creative Commons Namensnennung 4.0 International Lizenz
Fachbereich / Einrichtung:Medizinische Fakultät
Dokument erstellt am:04.03.2025
Dateien geändert am:04.03.2025
english
Benutzer
Status: Gast
Aktionen