Dokument: Standardization of a CT Protocol for Imaging Patients with Suspected COVID-19—A RACOON Project

Titel:Standardization of a CT Protocol for Imaging Patients with Suspected COVID-19—A RACOON Project
URL für Lesezeichen:https://docserv.uni-duesseldorf.de/servlets/DocumentServlet?id=68373
URN (NBN):urn:nbn:de:hbz:061-20250131-090855-2
Kollektion:Publikationen
Sprache:Englisch
Dokumententyp:Wissenschaftliche Texte » Artikel, Aufsatz
Medientyp:Text
Autoren: Steuwe, Andrea [Autor]
Kamp, Benedikt [Autor]
Afat, Saif [Autor]
Akinina, Alena [Autor]
Aludin, Schekeb [Autor]
Bas, Elif Gülsah [Autor]
Berger, Josephine [Autor]
Bohrer, Evelyn [Autor]
Brose, Alexander [Autor]
Büttner, Susanne Martina [Autor]
Dateien:
[Dateien anzeigen]Adobe PDF
[Details]1,56 MB in einer Datei
[ZIP-Datei erzeugen]
Dateien vom 31.01.2025 / geändert 31.01.2025
Stichwörter:COVID-19, radiation exposure, image quality, computed tomography
Beschreibung:CT protocols that diagnose COVID-19 vary in regard to the associated radiation exposure and the desired image quality (IQ). This study aims to evaluate CT protocols of hospitals participating in the RACOON (Radiological Cooperative Network) project, consolidating CT protocols to provide recommendations and strategies for future pandemics. In this retrospective study, CT acquisitions of COVID-19 patients scanned between March 2020 and October 2020 (RACOON phase 1) were included, and all non-contrast protocols were evaluated. For this purpose, CT protocol parameters, IQ ratings, radiation exposure (CTDIvol), and central patient diameters were sampled. Eventually, the data from 14 sites and 534 CT acquisitions were analyzed. IQ was rated good for 81% of the evaluated examinations. Motion, beam-hardening artefacts, or image noise were reasons for a suboptimal IQ. The tube potential ranged between 80 and 140 kVp, with the majority between 100 and 120 kVp. CTDIvol was 3.7 ± 3.4 mGy. Most healthcare facilities included did not have a specific non-contrast CT protocol. Furthermore, CT protocols for chest imaging varied in their settings and radiation exposure. In future, it will be necessary to make recommendations regarding the required IQ and protocol parameters for the majority of CT scanners to enable comparable IQ as well as radiation exposure for different sites but identical diagnostic questions.
Rechtliche Vermerke:Originalveröffentlichung:
Steuwe, A., Kamp, B., Afat, S., Akinina, A., Aludin, S., Bas, E. G., Berger, J., Bohrer, E., Brose, A., Büttner, S. M., Ehrengut, C., Gerwing, M., Grosu, S., Gussew, A., Güttler, F., Heinrich, A., Jiraskova, P., Kloth, C., Kottlors, J., … Valentin, B. (2024). Standardization of a CT Protocol for Imaging Patients with Suspected COVID-19—A RACOON Project. Bioengineering, 11(3), Article 207. https://doi.org/10.3390/bioengineering11030207
Lizenz:Creative Commons Lizenzvertrag
Dieses Werk ist lizenziert unter einer Creative Commons Namensnennung 4.0 International Lizenz
Fachbereich / Einrichtung:Medizinische Fakultät
Dokument erstellt am:31.01.2025